ResMed Ltd. recalls Astral 100 and Astral 150 ventilators

Recalled November 1, 2023 · FDA recall Z-0111-2024 · ResMed

Hazard
If ventilator is on internal battery, not intended to serve as a primary power source, low/critically low battery alarms will sound, but a fault leads to sudden power loss. If power fails, then ventilation stops and a Total Power Failure (TPF) alarm should sound, but it's powered by a supercapacitor, which degrades over time, which may cause TPF alarm to sound for less than 2 minutes or not at all · FDA Class I
Units
55,279 (US 16,634; OUS 38,645)
Sold at
Worldwide - US Nationwide distribution including in the states of CA, NJ, VA, IN, MI, WV, NY, WI, ND, IL, NC, FL, CO, TX, AL, LA, MD, OR, AZ, MN, WA, PA, SC, MA, ID, GA, SD, TN, NH, OH, MS, CT, IA, KY, UT, NE, MO, NV, AR, WY, RI, KS, VT, ME, OK, PR, MO, DC and the countries of Taiwan, Israel, Spain, Slovenia, Belgium, Italy, Australia, Guatemala, Argentina, Uruguay, Chile, France, Martinique, Germany, French Polynesia, Korea, Egypt, Colombia, United Kingdom, Jersey, Ecuador, Reunion, Greece, Brazil, Cyprus, Saudi Arabia, Czech Republic, Thailand, New Caledonia, Mauritius, Canada, Vietnam, Indonesia, Pakistan, Japan, Guadeloupe, South Africa, Iran, Hungary, Tunisia, Nepal, Iceland, New Zealand, Austria, Portugal, Malaysia, Norway, Netherlands, Sweden, Lebanon, Finland, Singapore, United Arab Emirates, Peru, Oman, Philippines, Bahrain, French Guiana, Mexico, Sri Lanka, India, Kuwait, Bangladesh, Kenya, Myanmar, Turkey, Bulgaria, Croatia, Hong Kong, Saint Martin, Qatar, Jordan, Romania, Maldives, Denmark, Poland.

Description

If ventilator is on internal battery, not intended to serve as a primary power source, low/critically low battery alarms will sound, but a fault leads to sudden power loss. If power fails, then ventilation stops and a Total Power Failure (TPF) alarm should sound, but it's powered by a supercapacitor, which degrades over time, which may cause TPF alarm to sound for less than 2 minutes or not at all

Product
Astral 100 and Astral 150 ventilators
Manufacturer(s)
ResMed Ltd.
Made in
Australia

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov