RAYSEARCH LABORATORIES AB recalls RayStation 4.0, 4.5, 4.7 and 5.0; Radiation Therapy Treatment Planning System Product Usage: RayStation is a softwar
Recalled December 14, 2016 · FDA recall Z-0721-2017 · RAYSEARCH LABORATORIES AB
Hazard
Software anomaly; an issue was found with the proton Pencil Beam Scanning (PBS) dose calculation in RayStation 4.0, 4.5, 4.7 and 5.0. For treatment plans with a combination of range shifter, large air gap and beams that enter the patient surface at an oblique angle, the dose calculation accuracy may be less than expected. · FDA Class II
Units
US: 66 units
Sold at
US Nationwide Distribution in the states of California, Florida, Illinois, Louisiana, Michigan, New Jersey, Tennessee, Texas, Washington
Description
Software anomaly; an issue was found with the proton Pencil Beam Scanning (PBS) dose calculation in RayStation 4.0, 4.5, 4.7 and 5.0. For treatment plans with a combination of range shifter, large air gap and beams that enter the patient surface at an oblique angle, the dose calculation accuracy may be less than expected.
Product
RayStation 4.0, 4.5, 4.7 and 5.0; Radiation Therapy Treatment Planning System Product Usage: RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs. The intended users of RayStation shall be clinically qualified radiation therapy staff trained in using the system.
Manufacturer(s)
RAYSEARCH LABORATORIES AB