Home / Recalls / Medical Devices
RAYSEARCH LABORATORIES AB recalls RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6, RayStation 7 and RayStation
Recalled January 2, 2019 · FDA recall Z-0679-2019 · RAYSEARCH LABORATORIES AB
Hazard
Isocenter shifts when using dose tracking in RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6, RayStation 7 and RayStation 8A. In these versions, isocenter shifts in dose tracking are incorrectly interpreted according to the DICOM patient coordinate system. · FDA Class II
Units
69
Sold at
AZ CA CO CT DE FL GA HI ID IL LA MA MD ME MI MO MS NC NJ NY OH OR PA RI TN TX VA VT WA WI
Description
Isocenter shifts when using dose tracking in RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6, RayStation 7 and RayStation 8A. In these versions, isocenter shifts in dose tracking are incorrectly interpreted according to the DICOM patient coordinate system.
Product
RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6, RayStation 7 and RayStation 8A
Manufacturer(s)
RAYSEARCH LABORATORIES AB
Made in
Sweden
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →