RecallWatchUSA

Home / Recalls / Medical Devices

RAYSEARCH LABORATORIES AB recalls Ray Station Radiation Therapy Treatment Planning System, Model Numbers 3.0, 3.5 and 4.0

Recalled November 20, 2013 · FDA recall Z-0281-2014 · RAYSEARCH LABORATORIES AB

Hazard
Ray Search Laboratories found an issue with the dose calculation for Elekta motorized wedges in Ray Station 3.0, 3.5 and 4.0. The error presents when a wedge is used for a beam and the wedge angle differs from 60 degrees. · FDA Class II
Units
195 units domestically
Sold at
US Distribution in the states of: California, Colorado, Delaware, Florida, Maryland, Missouri, New Jersey, New York, Ohio, Pennsylvania, Tennessee, and Virginia.

Description

Ray Search Laboratories found an issue with the dose calculation for Elekta motorized wedges in Ray Station 3.0, 3.5 and 4.0. The error presents when a wedge is used for a beam and the wedge angle differs from 60 degrees.

Product
Ray Station Radiation Therapy Treatment Planning System, Model Numbers 3.0, 3.5 and 4.0
Manufacturer(s)
RAYSEARCH LABORATORIES AB
Made in
Sweden

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

RAYSEARCH LABORATORIES AB recalls Ray Station Radiation Therapy Treatment Planning System, Model Numbers 3.0, 3.5 and 4.0 | RecallWatchUSA