Quidel Corporation recalls Lyra Direct SARS-CoV-2 Assay Emergency Use Authorization- Ref: M124. The Lyra Direct SARS-CoV-2 Assay is a real-time

Recalled September 22, 2021 · FDA recall Z-2477-2021 · Quidel

Hazard
Revised Instructions for Use: Potential for specimens from patients with high SARS-CoV-2 viral loads to be falsely interpreted as negative if target amplification occurs prior to a cycle-threshold (Ct) value of 5 when using the Thermo Fisher QuantStudio 7 Pro, the Applied Biosystems 7500 Fast Dx, the Applied Biosystems 7500, the Bio-Rad CFX96 Touch", the Roche LightCycler LC 480 II/ cobas z 480, or the Qiagen Rotor-Gene¿ MDx. · FDA Class II
Units
81,910 kits
Sold at
Worldwide distribution - US Nationwide distribution including Puerto Rico and the countries of Canada, Panama, Virgin Islands, Germany, Honduras.

Description

Revised Instructions for Use: Potential for specimens from patients with high SARS-CoV-2 viral loads to be falsely interpreted as negative if target amplification occurs prior to a cycle-threshold (Ct) value of 5 when using the Thermo Fisher QuantStudio 7 Pro, the Applied Biosystems 7500 Fast Dx, the Applied Biosystems 7500, the Bio-Rad CFX96 Touch", the Roche LightCycler LC 480 II/ cobas z 480, or the Qiagen Rotor-Gene¿ MDx.

Product
Lyra Direct SARS-CoV-2 Assay Emergency Use Authorization- Ref: M124. The Lyra Direct SARS-CoV-2 Assay is a real-time RT-PCR assay intended for the qualitative detection of nucleic acid from SARS-CoV-2 in nasal (NS), nasopharyngeal (NP), or oropharyngeal (OP) direct swab specimens from individuals suspected of COVID-19 by their healthcare provider
Manufacturer(s)
Quidel Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov