Philips North America Llc recalls Philips Hemodynamic Application, Model Number 722463, Software Version Number: R1.2.2 and R1.2.3. For visualization and

Recalled May 18, 2022 · FDA recall Z-1041-2022 · Philips North America

Hazard
When a Neonatal patient is admitted, the Non-Invasive Blood Pressure is not set to Neonatal mode on the Philips Xperflex Cardio FC2010 patient monitor. If the NIBP is not set to Neonatal, the cuff inflation pressure applied will be higher than 150 mmHg and the deflation pressure will be higher than 5 mmHg. An applied cuff inflation pressure higher than 150mmHg or a deflation pressure higher than 5 could lead to bruising, mmHg nerve damage or bone deformity for a neonate. · FDA Class II
Units
77
Sold at
Devices were distributed to the following US states: FL, ID, NJ, OK, SC and TX. Devices were distributed to the following foreign countries: France, Germany, India, Mexico, Netherlands, New Zealand, Norway, Poland, South Africa, and Switzerland.

Description

When a Neonatal patient is admitted, the Non-Invasive Blood Pressure is not set to Neonatal mode on the Philips Xperflex Cardio FC2010 patient monitor. If the NIBP is not set to Neonatal, the cuff inflation pressure applied will be higher than 150 mmHg and the deflation pressure will be higher than 5 mmHg. An applied cuff inflation pressure higher than 150mmHg or a deflation pressure higher than 5 could lead to bruising, mmHg nerve damage or bone deformity for a neonate.

Product
Philips Hemodynamic Application, Model Number 722463, Software Version Number: R1.2.2 and R1.2.3. For visualization and analysis of patient hemodynamic data in the Philips Interventional Hemodynamic System.
Manufacturer(s)
Philips North America Llc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov