Philips North America Llc recalls Philips CT systems labeled as the following: 1. Ingenuity CT 728326, Model Number: 728321. 2. Ingenuity CT 728326

Recalled October 1, 2025 · FDA recall Z-2600-2025 · Philips North America

Hazard
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement. · FDA Class II
Units
285 units
Sold at
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lebanon, Mauritius, Mexico, Netherlands, Norway, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, T¿rkiye, United Kingdom, Utd.Arab.Emir., Vietnam.

Description

The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.

Product
Philips CT systems labeled as the following: 1. Ingenuity CT 728326, Model Number: 728321. 2. Ingenuity CT 728326, Model Number: 728323. 3. Ingenuity CT, Model Number: 728326.
Manufacturer(s)
Philips North America Llc