Philips North America Llc recalls Philips Allura CV20 Interventional Fluoroscopic X-ray System products with a Laird chiller without an extended drip tray
Recalled March 16, 2022 · FDA recall Z-0725-2022 · Philips North America
Hazard
Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire. · FDA Class II
Units
1 unit
Sold at
US Nationwide Foreign: Argentina Austria Bangladesh Brazil Chile China Denmark Dominican Republic France Germany Greece India Indonesia Iran Israel Italy Japan Korea, Republic of Malaysia Mexico Netherlands Philippines Russian Federation Saudi Arabia South Africa Spain Sri Lanka Switzerland Taiwan Thailand Turkey United Kingdom Virgin Islands (U.S.)
Description
Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.
Product
Philips Allura CV20 Interventional Fluoroscopic X-ray System products with a Laird chiller without an extended drip tray: Model Number: 722031
Manufacturer(s)
Philips North America Llc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →