Home / Recalls / Medical Devices
Philips North America, LLC recalls Cardio Vascular-Allura Centron, system code 722400
Recalled March 25, 2020 · FDA recall Z-1494-2020 · Philips North America
Hazard
A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible. · FDA Class II
Units
87
Sold at
Nationwide domestic distribution. Foreign distribution worldwide.
Description
A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.
Product
Cardio Vascular-Allura Centron, system code 722400
Manufacturer(s)
Philips North America, LLC
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →