RecallWatchUSA

Home / Recalls / Medical Devices

Philips North America recalls IntelliVue Patient Monitor MX400 (Model Number 866060), MX430 (Model Number 866061), MX450 (Model Number 866062), MX500

Recalled October 2, 2024 · FDA recall Z-3249-2024 · Philips North America

Hazard
During a production process, Philips became aware of one IntelliVue power supply with a broken ground bolt upon disconnection of the ground cable from the equipotential ground connector. Loss of electrical grounding may negatively affect the devices electromagnetic immunity and emission. · FDA Class II
Units
545 devices
Sold at
US Nationwide. Global Distribution.

Description

During a production process, Philips became aware of one IntelliVue power supply with a broken ground bolt upon disconnection of the ground cable from the equipotential ground connector. Loss of electrical grounding may negatively affect the devices electromagnetic immunity and emission.

Product
IntelliVue Patient Monitor MX400 (Model Number 866060), MX430 (Model Number 866061), MX450 (Model Number 866062), MX500 (Model Number 866064), and MX550 (Model Number 866066)
Manufacturer(s)
Philips North America

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov