RecallWatchUSA

Home / Recalls / Medical Devices

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion system; Software Version Number: R3.1;

Recalled January 28, 2026 · FDA recall Z-1113-2026 · PHILIPS MEDICAL SYSTEMS NEDERLAND B.V

Hazard
Philips has identified two (2) software issues affecting device systems that may result in loss of imaging (X-ray) functionality and/or loss of motorized movement, and/or incorrect image content and/or loss of data. Issue 1 - System remains in continuous restart mode after the start-up. Issue 2 - Longitudinal Position Error Applicable only to systems with Poly-G3 frontal stand. · FDA Class II
Units
228 units (all OUS)
Sold at
International distribution in the countries of AE, AR, AT, AZ, BE, BG, BH, BR, BS, CH, CL, CO, CR, CY, CZ, DE, DK, DZ, EC, EG, ES, GB, GR, GT, ID, IE, IL, IN, IQ, IT, JE, JP, LB, MA, MO, NI, NL, NO, PE, PL, RO, SA, SE, SV, TR, UA, VE, VN, ZA;

Description

Philips has identified two (2) software issues affecting device systems that may result in loss of imaging (X-ray) functionality and/or loss of motorized movement, and/or incorrect image content and/or loss of data. Issue 1 - System remains in continuous restart mode after the start-up. Issue 2 - Longitudinal Position Error Applicable only to systems with Poly-G3 frontal stand.

Product
Philips Azurion system; Software Version Number: R3.1;
Manufacturer(s)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Made in
Netherlands

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov