RecallWatchUSA

Home / Recalls / Medical Devices

Olympus Corporation of the Americas recalls SOLTIVE SuperPulsed Laser Fibers Model Numbers TFL-FBX150S, TFL-FBX200S, TFL-FBX365S, TFL-FBX550S, TFL-FBX940S, TFL-FBX

Recalled January 24, 2024 · FDA recall Z-0748-2024 · Olympus Corporation of the Americas

Hazard
Olympus identified inconsistencies in the Instructions for Use regarding cleaning the fiber tip and fiber cleaving and stripping instructions. Following the current instructions presents the potential for the fiber to break during use, which may cause burns, procedure delays, prolonged surgery, and foreign body in patient. · FDA Class II
Units
82,200 total units (5 per box)
Sold at
Worldwide distribution - US Nationwide and the countries of Canada, Europe, Hong Kong, Australia, Japan, Singapore, and Latin America.

Description

Olympus identified inconsistencies in the Instructions for Use regarding cleaning the fiber tip and fiber cleaving and stripping instructions. Following the current instructions presents the potential for the fiber to break during use, which may cause burns, procedure delays, prolonged surgery, and foreign body in patient.

Product
SOLTIVE SuperPulsed Laser Fibers Model Numbers TFL-FBX150S, TFL-FBX200S, TFL-FBX365S, TFL-FBX550S, TFL-FBX940S, TFL-FBX150BS, TFL-FBX200BS
Manufacturer(s)
Olympus Corporation of the Americas

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov