RecallWatchUSA

Home / Recalls / Medical Devices

Novo Nordisk Inc recalls Dial-A-Dose Insulin Delivery Device (Pen-Injector) cartridge holders Product Usage: The NovoPen Echo¿ is a re-usable

Recalled December 13, 2017 · FDA recall Z-0206-2018 · Novo Nordisk

Hazard
A small number of batches may crack or break if exposed to certain chemicals, like certain cleaning agents. · FDA Class II
Units
16,557 units
Sold at
Nationwide distribution

Description

A small number of batches may crack or break if exposed to certain chemicals, like certain cleaning agents.

Product
Dial-A-Dose Insulin Delivery Device (Pen-Injector) cartridge holders Product Usage: The NovoPen Echo¿ is a re-usable pen injector for single patient use by diabetics for the self injection of a desired dose of insulin. The pen injector uses Penfill 3 mL cartridge of NovoLog¿, 100 units/mL (Ux100) [insulin aspart (rDNA origin) injection] and a single use detachable and disposable pen needle (supplied separately). The pen injector allows the user to dial the desired dose from 0.5 to 30 units in 0.5 increments.
Manufacturer(s)
Novo Nordisk Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov