Nihon Kohden America Inc recalls CNS-6200 Series Central Nurse Station and accessories. Model Number: CNS-6201 For cardiac and vital signs monitoring
Recalled July 9, 2014 · FDA recall Z-1979-2014 · Nihon Kohden America
Hazard
Software Version 02-26, when used with the Central Nursing Station (CNS) 6201, (PU-621 RA) may unexpectedly and without warning reboot, resulting in a period of approximately 3 minutes during which time the system does not display patient data and does not emit alarms based on that data should they be appropriate. · FDA Class II
Units
430
Sold at
Worldwide Distribution - US in the state of Oklahoma and in the countries of Bolivia, Chile, Germany, Japan, Korea, Taiwan, Yemen, and South Africa.
Description
Software Version 02-26, when used with the Central Nursing Station (CNS) 6201, (PU-621 RA) may unexpectedly and without warning reboot, resulting in a period of approximately 3 minutes during which time the system does not display patient data and does not emit alarms based on that data should they be appropriate.
Product
CNS-6200 Series Central Nurse Station and accessories. Model Number: CNS-6201 For cardiac and vital signs monitoring for multiple patients.
Manufacturer(s)
Nihon Kohden America Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →