RecallWatchUSA

Home / Recalls / Medical Devices

Micro-X Ltd. recalls MICRO-X Rover Mobile X-ray System, # MXU-RV35

Recalled December 18, 2024 · FDA recall Z-0664-2025 · Micro-X

Hazard
Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started. · FDA Class II
Units
5
Sold at
Worldwide and US Nationwide Distribution: NJ, PR, CA

Description

Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.

Product
MICRO-X Rover Mobile X-ray System, # MXU-RV35
Manufacturer(s)
Micro-X Ltd.
Made in
Australia

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov