Medtronic Perfusion Systems recalls RESERVOIR "Y" ADAPTER, (a) BT920 (b) BT926 Suction and Anticoagulation assemblies are indicated for use with an au

Recalled February 6, 2019 · FDA recall Z-0785-2019 · Medtronic Perfusion Systems

Hazard
Medtronic discovered that certain Y-connectors included in the suction lines have exhibited tearing and cracking. These defects could cause the Y-connectors to leak when in use, which could result in potential blood loss and a breach in the circuit. · FDA Class II
Units
1100 units
Sold at
Worldwide Distribution: US (nationwide) and countries of: Australia, Canada, France, Germany, Japan, and Sweden.

Description

Medtronic discovered that certain Y-connectors included in the suction lines have exhibited tearing and cracking. These defects could cause the Y-connectors to leak when in use, which could result in potential blood loss and a breach in the circuit.

Product
RESERVOIR "Y" ADAPTER, (a) BT920 (b) BT926 Suction and Anticoagulation assemblies are indicated for use with an autotransfusion system between the suction tip and reservoir.
Manufacturer(s)
Medtronic Perfusion Systems