Medtronic Perfusion Systems recalls Medtronic Perfusion Tubing Packs, Models TL7M64R3, TL7P17R1 & HY2W85R9, Sterilized using Ethylene Oxide. The Tubing P

Recalled November 18, 2015 · FDA recall Z-0268-2016 · Medtronic Perfusion Systems

Hazard
This recall is being initiated due to a potential breach of the Tubing Pack Header Bags manufactured in these select lots. The seal in question is between the Tyvek and the Poly side of the header bag. · FDA Class II
Units
496
Sold at
Nationwide Distribution including US: AL, IL, IN, and WI.

Description

This recall is being initiated due to a potential breach of the Tubing Pack Header Bags manufactured in these select lots. The seal in question is between the Tyvek and the Poly side of the header bag.

Product
Medtronic Perfusion Tubing Packs, Models TL7M64R3, TL7P17R1 & HY2W85R9, Sterilized using Ethylene Oxide. The Tubing Pack design and components are specified by the user, they are packaged with consideration of setup efficiency and respect for the environment, and they are intended for use in the extracorporeal circuit during cardiopulmonary bypass procedures. Medtronic Tubing Packs can contain components with Carmeda BioActive Surface, Trillium BioSurface, Balance BioSurface or no coating. Sterile barrier for these packs is provided either by a Tyvek lid seal on the tray or by a header bag.
Manufacturer(s)
Medtronic Perfusion Systems

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov