RecallWatchUSA

Home / Recalls / Medical Devices

Medtronic Neurosurgery recalls Medtronic Clip Gun Magazines are packaged in individually sealed pouches and distributed in boxes containing 5 individua

Recalled October 3, 2018 · FDA recall Z-3212-2018 · Medtronic Neurosurgery

Hazard
The packaging sterile barrier does not comply with the applicable packaging standard for terminally sterilized medical devices. While the Instructions for Use (IFU) state Do not use if package has been previously opened or damaged , it is possible that the seal breach may go undetected by the user's naked eye. · FDA Class II
Units
2616
Sold at
Domestic: AL, AZ, AR, CA, CO, DC, FL, GA, ID, IL, IN, IA, KS, LA, MD, MI, MN, MS, MO, NB, NV, NJ, NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WV, WI, and WY OUS: Australia, Germany, Kenya, Russian Federation, South Africa

Description

The packaging sterile barrier does not comply with the applicable packaging standard for terminally sterilized medical devices. While the Instructions for Use (IFU) state Do not use if package has been previously opened or damaged , it is possible that the seal breach may go undetected by the user's naked eye.

Product
Medtronic Clip Gun Magazines are packaged in individually sealed pouches and distributed in boxes containing 5 individually pouched Clip Gun Magazine. REF CM8902, Medtronic Clip Gun Magazine, Rx only, CE 0344, STERILE EO, Rx only, Single Use
Manufacturer(s)
Medtronic Neurosurgery

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov