Medtronic Neuromodulation recalls Medtronic Prostiva RF Therapy Generator. The Prostiva RF system is indicated for the treatment of symptoms due to uri
Recalled April 10, 2013 · FDA recall Z-1031-2013 · Medtronic Neuromodulation
Hazard
Medtronic has found through device testing that if the optional footswitch is not connected to the Model 8930 RF generator during use, an electrostatic discharge (i.e. static electricity) to the footswitch connector port may toggle the RF output on or off. If this were to occur, it could result in unintentionally starting or stopping the lesion process, depending on the current status of the equi · FDA Class II
Units
848 (696 US, 152 OUS)
Sold at
Nationwide Distribution including Puerto Rico and all States in continental USA except ID, MT, and NH.
Description
Medtronic has found through device testing that if the optional footswitch is not connected to the Model 8930 RF generator during use, an electrostatic discharge (i.e. static electricity) to the footswitch connector port may toggle the RF output on or off. If this were to occur, it could result in unintentionally starting or stopping the lesion process, depending on the current status of the equi
Product
Medtronic Prostiva RF Therapy Generator. The Prostiva RF system is indicated for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH) in men over the age of 50 with prostate sizes between 20 and 50 cm3.
Manufacturer(s)
Medtronic Neuromodulation
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →