RecallWatchUSA

Home / Recalls / Medical Devices

Medtronic MiniMed recalls Insulin Pump/Model: MiniMed 630G/ MMT-1715, MMT-1755, MMT-1754; MiniMed 670G/ MMT-1780, MMT-1781, MMT-1782, MMT-1760,

Recalled November 16, 2022 · FDA recall Z-0193-2023 · Medtronic MiniMed

Hazard
Medtronic identified a cybersecurity vulnerability in the MiniMed 600 series Insulin Pump Systems associated with the communication protocol that could allow unauthorized access to the pump system. This unauthorized access could be used to deliver too much or too little insulin through delivery of an unintended insulin bolus or because insulin delivery is slowed or stopped, which could lead to hypoglycemia or hyperglycemia. The Remote Bolus feature on insulin pumps should be turned off to prevent the unauthorized access. · FDA Class II
Units
399651
Sold at
US: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY. OUS: Andorra, Australia, Austria, Barbados, Belgium, Bermuda, Bolivia, Botswana, Canada, Canary Islands, Cayman Islands, Chile, Czech Republic, Denmark, Dominican Rep., El Salvador, Faroe Islands, Finland, France, Germany, Gibraltar, Greece, Guatemala, Hungary, Ireland, Israel, Italy, Kenya, Liechtenstein, Luxembourg, Mauritius, Mexico, Namibia, Netherlands, New Zealand, Norway, Paraguay, Peru, Poland, Portugal, Romania, Slovakia, South Africa, Spain, Sweden, Switzerland, Trinidad and Tobago, Turkey, Uganda, United Kingdom, Venezuela

Description

Medtronic identified a cybersecurity vulnerability in the MiniMed 600 series Insulin Pump Systems associated with the communication protocol that could allow unauthorized access to the pump system. This unauthorized access could be used to deliver too much or too little insulin through delivery of an unintended insulin bolus or because insulin delivery is slowed or stopped, which could lead to hypoglycemia or hyperglycemia. The Remote Bolus feature on insulin pumps should be turned off to prevent the unauthorized access.

Product
Insulin Pump/Model: MiniMed 630G/ MMT-1715, MMT-1755, MMT-1754; MiniMed 670G/ MMT-1780, MMT-1781, MMT-1782, MMT-1760, MMT-1761, MMT-1762, MMT-1740, MMT-1741, MMT-1742
Manufacturer(s)
Medtronic MiniMed

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Medtronic MiniMed recalls Insulin Pump/Model: MiniMed 630G/ MMT-1715, MMT-1755, MMT-1754; MiniMed 670G/ MMT-1780, MMT-1781, MMT-1782, MMT-1760, | RecallWatchUSA