Medline Industries, LP recalls NAMIC Angiographic Control Syringe, labeled as follows: 1. SYRINGE,7ML,TR/FR,RA,W/RES,-,P/G, Medline SKU 70083007;

Recalled April 15, 2026 · FDA recall Z-1713-2026 · Medline Industries

Hazard
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold. · FDA Class I
Units
192690 units
Sold at
Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).

Description

Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

Product
NAMIC Angiographic Control Syringe, labeled as follows: 1. SYRINGE,7ML,TR/FR,RA,W/RES,-,P/G, Medline SKU 70083007; 2. SYRINGE,8ML,TR/FR,RA,W/RES,-,P/G, Medline SKU 70084007; 3. MTS,SYRINGE,8ML,TR/FR,RA,NO,RES,-,P/G, Medline SKU 70084017; 4. SYRINGE,7,ML,TR/FR,RA,W/RES,OEM,PG, Medline SKU 80083007; 5. SYRINGE,8,ML,TR/FR,RA,W/RES,OEM,PG, Medline SKU 80084007; 6. MTO,SYRINGE,8,ML,TR/FR,RA,NO,RES,OEM,PG, Medline SKU 80084017.
Manufacturer(s)
Medline Industries, LP

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov