Home / Recalls / Medical Devices
Medline Industries, LP recalls Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: DR HOLCOMB FACIAL P
Recalled January 21, 2026 · FDA recall Z-1054-2026 · Medline Industries
Hazard
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions. · FDA Class II
Units
16 units
Sold at
US Nationwide distribution.
Description
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
Product
Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: DR HOLCOMB FACIAL PACK, Medline kit SKU DYNJ54488A
Manufacturer(s)
Medline Industries, LP
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →