RecallWatchUSA

Home / Recalls / Medical Devices

Medline Industries, LP recalls Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; ENT DYNJ909128I; LARYNGOSCOPY - BRON

Recalled November 19, 2025 · FDA recall Z-0445-2026 · Medline Industries

Hazard
Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment. · FDA Class I
Units
96,406 total units
Sold at
International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;

Description

Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.

Product
Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; ENT DYNJ909128I; LARYNGOSCOPY - BRONCHOSCOPY DYNJ909114D
Manufacturer(s)
Medline Industries, LP

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov