RecallWatchUSA

Home / Recalls / Medical Devices

Mckesson Medical-Surgical Inc. Corporate Office recalls Strep test: a. McKesson Medical-Surgical McKesson Rapid Test Kit McKesson Consult Infectious Disease Immunoassay Strep

Recalled August 31, 2022 · FDA recall Z-1598-2022 · Mckesson Medical-Surgical Inc. Corporate Office

Hazard
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21. · FDA Class II
Units
a. 214 kits b. 1cs c. 53 kits d. 4 bx e. 1 KT
Sold at
US Nationwide

Description

Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.

Product
Strep test: a. McKesson Medical-Surgical McKesson Rapid Test Kit McKesson Consult Infectious Disease Immunoassay Strep A Test Throat / Tonsil Saliva Sample 25 Tests, Model Number: 5003. b. QUIDEL Sofia TEST KIT, STREP A+ SOFIA (25/KT 12KT/CS) Model Number: 20274. c. Alere SAN DIEGO INC McKesson TEST, KIT STREP A DIPSTICK WAIVED (50/KT) Exempt Catalog #4999. d. In-Line Control Swab Quickvue In-Line Strep A Test Positive Level / Negative Level 12 Swabs (IVD) Catalog # 00345. e. (added 11/30/22) Abbott Rapid Test Kit Acceava Infectious Disease Immunoassay Strep A Test Throat Swab Sample 50 Tests, Model Number: 92001
Manufacturer(s)
Mckesson Medical-Surgical Inc. Corporate Office

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov