Intuitive Surgical, Inc. recalls daVinci XI Sureform 60 reload, 60 mm Green 6- Row, 4.6 mm, REF 48360G-08, UDI: 00886874113844, Rx Only, Sterile EO - Pr

Recalled May 13, 2020 · FDA recall Z-1890-2020 · Intuitive Surgical

Hazard
The firm became aware of a manufacturing variation in which black and green reloads in select Stapler Reloads may not deploy individual adjacent staples, which could result in an incomplete staple line. Failure of a incomplete staple like could potentially lead to an air leak or an anastomotic leak which may require an additional procedural intervention. · FDA Class II
Units
22,752 reloads
Sold at
Worldwide distribution - US Nationwide including in the states of AL, AK, AZ, AR, CA, CO, CT, DE, D.C., FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WV, WI and the countries of Australia, Czech Republic, Denmark, Finland, France, Germany, Italy, Japan, Luxembourg, Netherlands, Norway, South Korea, Spain, Switzerland, United Kingdom.

Description

The firm became aware of a manufacturing variation in which black and green reloads in select Stapler Reloads may not deploy individual adjacent staples, which could result in an incomplete staple line. Failure of a incomplete staple like could potentially lead to an air leak or an anastomotic leak which may require an additional procedural intervention.

Product
daVinci XI Sureform 60 reload, 60 mm Green 6- Row, 4.6 mm, REF 48360G-08, UDI: 00886874113844, Rx Only, Sterile EO - Product Usage: are intended to be used with da Vinci Surgical Systems for resection, transection, and, or creation of anastomoses in General, Thoracic, Gynecologic, Urologic, and Pediatric surgery.
Manufacturer(s)
Intuitive Surgical, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Intuitive Surgical, Inc. recalls daVinci XI Sureform 60 reload, 60 mm Green 6- Row, 4.6 mm, REF 48360G-08, UDI: 00886874113844, Rx Only, Sterile EO - Pr | RecallWatchUSA