RecallWatchUSA

Home / Recalls / Medical Devices

ICU Medical, Inc. recalls Spinning Spiros, Closed Male Luer, REF: 060-CH2000S; SURPLUG ChemoAccess Closed Male Luer, REF: 066-CH2000S; SURPLUG Che

Recalled December 30, 2020 · FDA recall Z-0700-2021 · ICU Medical

Hazard
Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line. · FDA Class II
Units
33900
Sold at
Worldwide distribution - U.S. Nationwide distribution including in the states of FL, MD, LA, PA, IN, TX, NY, MO, VA, OH, GA, NC, IL, CA, AZ, AK, NE, DE, WI, CO, UT, ND, MS, MA, MN, WA, SC, OR, KS, MI, NH, NJ, KY, TN, ID, AR, CT, OK, AL, SD, ME, IA, VT, WY, DC, NV, RI, MT, NM, HI, WV and the countries of Italy, Belgium, United Kingdom, Canada, Germany, Argentina, Australia, Poland, Greece, Switzerland, Israel, France, Finland, Spain, Malaysia, Taiwan, Ireland, Luxembourg, Hong Kong , Brazil, Netherlands, Austria, Sweden, Turkey, Singapore, Slovakia, Croatia, South Africa, Romania, Thailand.

Description

Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.

Product
Spinning Spiros, Closed Male Luer, REF: 060-CH2000S; SURPLUG ChemoAccess Closed Male Luer, REF: 066-CH2000S; SURPLUG ChemoAccess 5" (13cm) Bag Spike Adaptor with Closed Male Luer, REF: 066-CH3034; 35" (89 cm) Appx 3.9 ml, 10 Drop Admin Set w/Integrated Clave Drip Chamber, Clave, Spiros w/Red Cap, Bag Hanger, REF: CH3762; 35" (89 cm) Appx 6.6 ml, 10 Drop Admin Set w/Integrated Clave Drip Chamber, 0.2 Micron Low Protein Binding Filter, Clave, Spiros w/Red Cap, Bag Hanger, REF: CH3763
Manufacturer(s)
ICU Medical, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov