Home / Recalls / Medical Devices
ICU Medical Inc. recalls Hotline Fluid Warming Set, single use device, is designed to warm blood and intravenous fluids and deliver them to the p
Recalled July 29, 2026 · FDA recall Z-2759-2026 · ICU Medical
Hazard
ICU Medical became aware of an increased likelihood of leakage in specific Lots of L-70 HOTLINE Fluid Warming Sets due to potentially inadequate solvent application at the tubing to return connector bond. · FDA Class II
Units
39,750 units
Sold at
Worldwide - US Nationwide distribution.
Description
ICU Medical became aware of an increased likelihood of leakage in specific Lots of L-70 HOTLINE Fluid Warming Sets due to potentially inadequate solvent application at the tubing to return connector bond.
Product
Hotline Fluid Warming Set, single use device, is designed to warm blood and intravenous fluids and deliver them to the patient's intravenous access site at normothermic temperatures under gravity flow conditions, List Number L-70
Manufacturer(s)
ICU Medical Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →