Hazard
Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line. · FDA Class II
Units
55000
Sold at
Worldwide distribution - U.S. Nationwide distribution including in the states of FL, MD, LA, PA, IN, TX, NY, MO, VA, OH, GA, NC, IL, CA, AZ, AK, NE, DE, WI, CO, UT, ND, MS, MA, MN, WA, SC, OR, KS, MI, NH, NJ, KY, TN, ID, AR, CT, OK, AL, SD, ME, IA, VT, WY, DC, NV, RI, MT, NM, HI, WV and the countries of Italy, Belgium, United Kingdom, Canada, Germany, Argentina, Australia, Poland, Greece, Switzerland, Israel, France, Finland, Spain, Malaysia, Taiwan, Ireland, Luxembourg, Hong Kong , Brazil, Netherlands, Austria, Sweden, Turkey, Singapore, Slovakia, Croatia, South Africa, Romania, Thailand.

Description

Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.

Product
5 IN (13 cm) Appx 1.5 ml, Bag Spike Adapter, Spiros, REF: 20123-01; 75" (191 cm) Appx 9.1 ml, 20 Drop Admin Set w/ Integrated Clave Drip Chamber, Clave, Spiros, REF: CH3035; 40" (102 cm) Appx 4.5 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, Spiros, Hanger, REF: CH3135; 14" (36 cm) Appx 4.8 ml, Bifuse Add-On Set w/Spiros w/Red Cap, Bag Spike, 3 Clamps (Blue, 2 White), Vented Cap, Drop-In Spiros w/Red Cap, REF: CH3156; 114" Bifuse Primary Set w/2 Spiros, Bag Spike, 2 20 Drop In-line Drip Chamber, Clave, Bag Hanger, REF: CH3240; 18 IN (46 cm) Appx 2.9 ml, Add-On Set w/Spiros, Clave Spike, Clamp, Cap, REF: CH3290; 27" (69 cm) Appx 5.2 ml, Bifuse Add-On Set w/15 Drop Drip Chamber, Spiros w/Red Cap, Check Valve, Vented Cap, Bag Hanger, REF: CH3294; Secondary IV Oncology Kit, REF: CH3322; 30 IN (76 cm) Appx 1.8 ml) Ext Set w/2 Spiros, 2 Red Caps, Clamp, REF: CH3327; 40" (102 cm) Appx 4.6 ml, Admin Set, 2 Spiros w/2 Red Caps, 20 Drop In-Line Drip Chamber, 2 Bag Hangers, REF: CH3373; Oncology Kit w/Spiros, Red Cap, Bag Spike w/Clave Additive Port, REF: CH3397; 45" (114 cm) Appx 5.8 ml, Ext Set w/MicroClave, 2 Clave, Spiros w/Red Cap, 2 Clamps, REF: CH3403; Oncology Kit w/20" (51 cm) Ext Set w/MicroClave, Graduated Connector, Transfer Set w/Clave Additive Port, Spiros, Red Cap, 13mm Vial Spike w/Clave, REF: CH3507; Oncology Kit w/Clave Universal Vented Vial Spike, Spinning Spiros w/Red Cap, REF: CH3546; 22" (56 cm) Add-On 150 ml Burette Set (MicroClave, No Shut-Off), w/Spiros, 2 Clamps, Vented Cap, REF: CH3554; 4.5 IN (11 cm) Appx 0.43 ml, Ext Set, Spiros w/Red Cap, REF: CH3559; Oncology Kit w/Spinning Spiros, C-Clip, Red Cap, REF: CH3581; 5 IN (13 cm) Appx 0.57 ml, Ext Set, Anti-Siphon Valve, Spiros, REF: CH3582; Oncology Kit w/Spinning Spiros w/Red Cap, C-Clip; Priming Cap, REF: CH3589; 77" (196 cm) Appx. 9.4 ml, Admin Set w/2 Spiros w/Red Cap, 20 Drop In-Line Drip Chamber, Bag Hanger, REF: CH3639; 22" (56 cm) Appx 5.6ml, Ext Set, Trifuse w/4 Clave Clear, Spiros w/Red Cap, 1.2 Micron Filter, 4 Clamps, REF: CH3770; 6.5 IN (17 cm) Appx 0.78 ml, Ext Set w/Clave Clear, Clave 4-Way Stopcock, Spiros, Red Cap, REF: CH3771; 47" (119 cm) 50 ml Diluent Set for Channel 2 w/20 Drop In-Line Drip Chamber, 2 Spiros, REF: CH4001; 47" (119 cm) 50 ml Diluent Set For Channel 2 w/Bag Spike w/MicroClave Additive Port, Spiros, REF: CH4009; 30" (76 cm) Appx 5.9 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, 0.2 Micron Filter, Spiros, Hanger, REF: Z7100; 30" (76 cm) 50 ml Diluent Set for Channel 2 w/Spiros, REF: CH4006; 60 IN(152cm)APPX 0.45ml,SMALLBORE EXT,MicroCLAVE,Spiros, REF: Z7040; 6" (15 cm) Appx 0.55 ml, Smallbore Trifuse Ext Set w/3 MicroClave, 3 Clamps, Spiros, REF: Z7151
Manufacturer(s)
ICU Medical, Inc.