RecallWatchUSA

Home / Recalls / Medical Devices

Howmedica Osteonics Corp. recalls The Stryker CranialMask Tracker is a single-use device consisting of a flexible printed circuit board (PCB) with 31 infr

Recalled December 31, 2025 · FDA recall Z-0963-2026 · Howmedica Osteonics

Hazard
When the device is activated during surgery, the device software issues an error message "Instrument Face Mask (6001385000000) is not valid for this application" resulting in an inability to use the device. · FDA Class II
Units
320 units
Sold at
Worldwide distribution - US Nationwide and the countries of Canada, Germany, Japan, South Korea.

Description

When the device is activated during surgery, the device software issues an error message "Instrument Face Mask (6001385000000) is not valid for this application" resulting in an inability to use the device.

Product
The Stryker CranialMask Tracker is a single-use device consisting of a flexible printed circuit board (PCB) with 31 infrared (IR) light-emitting diodes (LEDs) as well as an electronics control unit for the LEDs which includes its own battery power supply. The measured LED positions can be used in one of three different modes of operation: Patient tracking, Registration using surface matching, and Automatic intra-operative mask registration.
Manufacturer(s)
Howmedica Osteonics Corp.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov