Hill-Rom, Inc. recalls Advanta 2 Bed. The Advanta 2 Bed is intended for low to moderate acuity patients in the medical/surgical area of the

Recalled July 4, 2012 · FDA recall Z-1864-2012 · Hill-Rom

Hazard
During an evaluation of Advanta 2 siderail samples returned from the field it was found that the screws used to attach the siderail were being stripped from the plastic cavity of the siderail. · FDA Class II
Units
7843 devices
Sold at
Worldwide Distribution-USA (nationwide) and the countries of Canada, France, and Singapore.

Description

During an evaluation of Advanta 2 siderail samples returned from the field it was found that the screws used to attach the siderail were being stripped from the plastic cavity of the siderail.

Product
Advanta 2 Bed. The Advanta 2 Bed is intended for low to moderate acuity patients in the medical/surgical area of the hospital.
Manufacturer(s)
Hill-Rom, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov