Home / Recalls / Medical Devices
GE Medical Systems Information Technologies Inc recalls GE Healthcare MAC VU360, REF# part number 2030361-023; Electrocardiograph
Recalled August 12, 2026 · FDA recall Z-2876-2026 · GE Medical Systems Information Technologies
Hazard
Software issue issue with the MAC VU360 v3.0 system. After selecting the order refresh icon, an open preview or review screen might fail to close automatically after 2 minutes of inactivity. As a result, data from a previous patient may remain on the screen and could be incorrectly associated with a subsequent patient. · FDA Class II
Units
163 units
Sold at
Worldwide distribution - US Nationwide and the countries of Belgium, Hong Kong, Netherlands, Sweden, UK.
Description
Software issue issue with the MAC VU360 v3.0 system. After selecting the order refresh icon, an open preview or review screen might fail to close automatically after 2 minutes of inactivity. As a result, data from a previous patient may remain on the screen and could be incorrectly associated with a subsequent patient.
Product
GE Healthcare MAC VU360, REF# part number 2030361-023; Electrocardiograph
Manufacturer(s)
GE Medical Systems Information Technologies Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →