GE Medical Systems China Co., Ltd. recalls GE 9100c NXT anesthesia systems

Recalled September 5, 2018 · FDA recall Z-2956-2018 · GE Medical Systems China

Hazard
On systems equipped with the Auxiliary Common Gas Outlet (ACGO), there is a potential for the auxiliary gas switch to unexpectedly switch from the ACGO position back to the circle position. On systems equipped with the Aux O2 + Air function the switch could unexpectedly switch from the Aux O2 + Air position back to the circle position. These situations could lead to hypoxia. · FDA Class II
Units
401 units
Sold at
Worldwide

Description

On systems equipped with the Auxiliary Common Gas Outlet (ACGO), there is a potential for the auxiliary gas switch to unexpectedly switch from the ACGO position back to the circle position. On systems equipped with the Aux O2 + Air function the switch could unexpectedly switch from the Aux O2 + Air position back to the circle position. These situations could lead to hypoxia.

Product
GE 9100c NXT anesthesia systems
Manufacturer(s)
GE Medical Systems China Co., Ltd.
Made in
China