Home / Recalls / Medical Devices
Fresenius Medical Care Holdings, Inc. recalls AquaC UNO H [120V, 60Hz, 1000VA, Fresenius US-Version: 24-10VA-1 Vivonic Part Number: G02000970-US
Recalled August 3, 2022 · FDA recall Z-1464-2022 · Fresenius Medical Care Holdings
Hazard
Internal ethernet cable may be misaligned, the cable jacket (insulation) could become compromised due to high temperature and result in the internal ethernet cable conducting electrical current to the ethernet port on the outside of the device), potentially harming service technicians and/or other authorized operators. If the ethernet port on the device is touched or manipulated, there is a risk of electrical shock · FDA Class II
Units
318 units
Sold at
US Nationwide distribution.
Description
Internal ethernet cable may be misaligned, the cable jacket (insulation) could become compromised due to high temperature and result in the internal ethernet cable conducting electrical current to the ethernet port on the outside of the device), potentially harming service technicians and/or other authorized operators. If the ethernet port on the device is touched or manipulated, there is a risk of electrical shock
Product
AquaC UNO H [120V, 60Hz, 1000VA, Fresenius US-Version: 24-10VA-1 Vivonic Part Number: G02000970-US
Manufacturer(s)
Fresenius Medical Care Holdings, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →