RecallWatchUSA

Home / Recalls / Medical Devices

Fresenius Kabi USA, LLC recalls Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004

Recalled June 10, 2026 · FDA recall Z-2224-2026 · Fresenius Kabi USA

Hazard
Fresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in service. More specifically, drops or jarring on a hard surface can cause the touchscreen input cable to loosen, resulting in unintended touchscreen behavior on the LVP, including random screen touches and/or lack of screen responsiveness to user input. Under the reported failure condition, unintended touch inputs may register at arbitrary screen locations without user interaction. These inputs occur intermittently and without a predictable pattern. · FDA Class I
Units
18,444
Sold at
Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI;

Description

Fresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in service. More specifically, drops or jarring on a hard surface can cause the touchscreen input cable to loosen, resulting in unintended touchscreen behavior on the LVP, including random screen touches and/or lack of screen responsiveness to user input. Under the reported failure condition, unintended touch inputs may register at arbitrary screen locations without user interaction. These inputs occur intermittently and without a predictable pattern.

Product
Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004
Manufacturer(s)
Fresenius Kabi USA, LLC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov