Recalled February 20, 2019 · FDA recall Z-0821-2019 · Elekta
Hazard
There is a potential that Wedge IDs were not included in the DICOM RT PLAN sent from the TPS, which will cause the field in MOSAIQ to be created with no wedge. Treatment of a field without the planned wedge will receive more dose than the treatment plan indicates. · FDA Class II
Units
3,950 units total (1,747 US)
Sold at
U.S. Nationwide distribution to the following states AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the District of Columbia and Puerto Rico.
Description
There is a potential that Wedge IDs were not included in the DICOM RT PLAN sent from the TPS, which will cause the field in MOSAIQ to be created with no wedge. Treatment of a field without the planned wedge will receive more dose than the treatment plan indicates.
Product
MOSAIQ¿ Oncology Information System and Sequencer MOSAIQ is an oncology information system used to manage workflows for treatment planning and delivery. It supports information flow among healthcare facility personnel and can be used wherever radiotherapy and/or chemotherapy are prescribed.
Manufacturer(s)
Elekta, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →