RecallWatchUSA

Home / Recalls / Medical Devices

Elekta, Inc. recalls ELEKTA LIMITED Elekta Infinity, REF XRT 0511, XRT 0521, XRT 0531, XRT 0541; Medical Charged-Particle Radiation Therapy S

Recalled December 13, 2023 · FDA recall Z-0446-2024 · Elekta

Hazard
Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension. · FDA Class II
Units
291 units
Sold at
worldwide

Description

Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension.

Product
ELEKTA LIMITED Elekta Infinity, REF XRT 0511, XRT 0521, XRT 0531, XRT 0541; Medical Charged-Particle Radiation Therapy System Accelerator, Linear
Manufacturer(s)
Elekta, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Elekta, Inc. recalls ELEKTA LIMITED Elekta Infinity, REF XRT 0511, XRT 0521, XRT 0531, XRT 0541; Medical Charged-Particle Radiation Therapy S | RecallWatchUSA