Home / Recalls / Medical Devices
Diversatek Healthcare recalls Diversatek Healthcare Viper¿ 3-Stage Fixed Wire Balloon Dilator, 6mm-7mm-8mm, REF 1204-08
Recalled April 29, 2026 · FDA recall Z-1938-2026 · Diversatek Healthcare
Hazard
The catheter inflation tag attached to the device may contain an incorrect part number with incorrect balloon diameter sizes and inflation pressures for each stage. · FDA Class II
Units
15 units
Sold at
US: MO, CA, FL, TX, AZ, PA, RI, NC, IL, MS, KS, NY, OH, VA, LA, IA, MI, NJ, AL, Brazil, Belgium
Description
The catheter inflation tag attached to the device may contain an incorrect part number with incorrect balloon diameter sizes and inflation pressures for each stage.
Product
Diversatek Healthcare Viper¿ 3-Stage Fixed Wire Balloon Dilator, 6mm-7mm-8mm, REF 1204-08
Manufacturer(s)
Diversatek Healthcare
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →