RecallWatchUSA

Home / Recalls / Medical Devices

Dexcom, Inc. recalls Dexcom G7 Glucose Receiver, SKUs (part number): STK-AT-011(MT26403-0), STK-AT-012 (MT26403-0), STK-AT-013 (MT26403-0),

Recalled June 25, 2025 · FDA recall Z-1944-2025 · Dexcom

Hazard
Defective foam or an assembly error may cause the receiver speaker to lose contact with the printed circuit board, leading to missed audible alerts for low or high blood glucose values. A missed audible alert for low or high blood glucose values could lead to untreated hypo or hyperglycemia which can cause seizures, vomiting, loss of consciousness, or death. · FDA Class I
Units
602,445 units
Sold at
worldwide distribution - US Nationwide and the countries of Andorra, Argentina Australia, Austria, Bahrain, Belgium, Bulgaria, Canada, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Greece, Hong Kong, Ireland, Israel, Italy, Japan, Korea, Latvia, Liechtenstein, Lithuania, Luxembourg, Malta, Netherlands, New Zealand, Norway, Oman, Poland, Romania, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, United Kingdom.

Description

Defective foam or an assembly error may cause the receiver speaker to lose contact with the printed circuit board, leading to missed audible alerts for low or high blood glucose values. A missed audible alert for low or high blood glucose values could lead to untreated hypo or hyperglycemia which can cause seizures, vomiting, loss of consciousness, or death.

Product
Dexcom G7 Glucose Receiver, SKUs (part number): STK-AT-011(MT26403-0), STK-AT-012 (MT26403-0), STK-AT-013 (MT26403-0), STK-GT-001 (MT26403-1), STK-GT-008 (MT26403-5), STK-GT-013 (MT26403-1) , STK-GT-019 (MT26403-4) , STK-GT-100 (MT26403-2), STK-GT-109 (MT26403-2) , STK-GT-113 (MT26403-3)
Manufacturer(s)
Dexcom, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov