RecallWatchUSA

Home / Recalls / Medical Devices

DeRoyal Industries Inc recalls DeRoyal THE RESOURCE GROUP, CRANI PACK, REF 89-9211.07 custom surgical pack

Recalled July 24, 2019 · FDA recall Z-1981-2019 · DeRoyal Industries

Hazard
Custom surgical packs are being recalled because it contains Cardinal Health Blunt Cannulas which were recalled due to a manufacturing defect in the cartridge component which could compromise the sterility barrier of the product. · FDA Class II
Units
108 units
Sold at
US Distribution to: OK

Description

Custom surgical packs are being recalled because it contains Cardinal Health Blunt Cannulas which were recalled due to a manufacturing defect in the cartridge component which could compromise the sterility barrier of the product.

Product
DeRoyal THE RESOURCE GROUP, CRANI PACK, REF 89-9211.07 custom surgical pack
Manufacturer(s)
DeRoyal Industries Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov