Home / Recalls / Medical Devices
DeRoyal Industries Inc recalls DeRoyal limb holders: (1) DeRoyal Limb Holder, Wrist, Fixed, Quick Release, REF M8124; (2) DeRoyal Limb Holder, Ankl
Recalled August 31, 2022 · FDA recall Z-1562-2022 · DeRoyal Industries
Hazard
The webbing strap was laced improperly through the teeth of the buckle allowing the webbing strap to slide and was, therefore, not secure. · FDA Class II
Units
36 eaches and 7 pairs
Sold at
US Nationwide distribution in the states of FL, IN, NV, RI, TN, TX, and WV.
Description
The webbing strap was laced improperly through the teeth of the buckle allowing the webbing strap to slide and was, therefore, not secure.
Product
DeRoyal limb holders: (1) DeRoyal Limb Holder, Wrist, Fixed, Quick Release, REF M8124; (2) DeRoyal Limb Holder, Ankle, Fixed, Quick Release, REF M8125; (3) DeRoyal Security Cuffs, Wrist and Ankle, Universal, REF M8136; and (4) DeRoyal Security Cuffs, Heavy Duty Vertical Psychiatric, University, REF M8137.
Manufacturer(s)
DeRoyal Industries Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →