DATEX--OHMEDA, INC. recalls GE Healthcare Giraffe Omnibed Carestation CS1, combination infant incubator and infant warmer. Model Numbers: 1) 208

Recalled November 6, 2024 · FDA recall Z-0140-2025 · DATEX--OHMEDA

Hazard
Prior to current established manufacturing practices, higher levels of airborne formaldehyde exposure from newly manufactured incubators could be seen during clinical use that would rapidly decrease over a week. Affected customers were notified in June 2023 to run incubators for a week to potentially off gas any residual airborne formaldehyde. The general scientific consensus is that sensory irritation of the eyes is the most sensitive endpoint with respect to exposure to airborne formaldehyde and is the adverse effect clinically experienced at the lowest concentration. · FDA Class I
Units
769 units
Sold at
worldwide

Description

Prior to current established manufacturing practices, higher levels of airborne formaldehyde exposure from newly manufactured incubators could be seen during clinical use that would rapidly decrease over a week. Affected customers were notified in June 2023 to run incubators for a week to potentially off gas any residual airborne formaldehyde. The general scientific consensus is that sensory irritation of the eyes is the most sensitive endpoint with respect to exposure to airborne formaldehyde and is the adverse effect clinically experienced at the lowest concentration.

Product
GE Healthcare Giraffe Omnibed Carestation CS1, combination infant incubator and infant warmer. Model Numbers: 1) 2082844-001-01040418; 2) 2082844-001-01046784; 3) 2082844-001-01052324; 4) 2082844-001-01052572; 5) 2082844-001-01055337; 6) 2082844-001-01065693; 7) 2082844-001-01069008; 8) 2082844-001-01073750; 9) 2082844-001-01086553; 10) 2082844-001-01089393; 11) 2082844-001-01090414; 12) 2082844-001-01092632; 13) 2082844-001-01111482; 14) 2082844-001-01115741; 15) 2082844-001-01122391; 16) 2082844-001-01128579; 17) 2082844-001-01128725; 18) 2082844-001-01131573; 19) 2082844-001-01140642; 20) 2082844-001-01165870; 21) 2082844-001-100397; 22) 2082844-001-100676; 23) 2082844-001-100740; 24) 2082844-001-100743; 25) 2082844-001-100744; 26) 2082844-001-101075; 27) 2082844-001-101149; 28) 2082844-001-101199; 29) 2082844-001-101202; 30) 2082844-001-101203; 31) 2082844-001-101356; 32) 2082844-001-101366; 33) 2082844-001-101656; 34) 2082844-001-102136; 35) 2082844-001-102589; 36) 2082844-001-102858; 37) 2082844-001-526493; 38) 2082844-001-533346; 39) 2082844-001-570904; 40) 2082844-001-615985; 41) 2082844-001-616544; 42) 2082844-001-674814; 43) 2082844-001-701655; 44) 2082844-001-704407; 45) 2082844-001-709042; 46) 2082844-001-857748; 47) 2082844-001-888278; 48) 2082844-001-941288; 49) 2082844-001-965539; 50) 2082844-001-978713; 51) 2082844-001-982099; 52) 2082844-001-982310; 53) 2082844-001-982985; 54) 2082844-001-982988; 55) 2082844-001-983509; 56) 2082844-001-983643; 57) 2082844-001-983679; 58) 2082844-001-984096; 59) 2082844-001-984948; 60) 2082844-001-987480; 61) 2082844-001-987855; 62) 2082844-001-993967; 63) 2082844-001-994441; 64) 2082844-001-995782; 65) 2082844-001-998377; Radiant infant warmer
Manufacturer(s)
DATEX--OHMEDA, INC.