Home / Recalls / Medical Devices
Covidien LLC recalls HALO Flex Energy Generators Manufactured by B¿RRX Medical (now Covidien) 540 Oakmead Parkway Sunnyvale, CA 95085.
Recalled September 26, 2012 · FDA recall Z-2379-2012 · Covidien
Hazard
Potential failure of HALO Energy Generator to enter "stand by" mode when initially powered on; and therefore can not be used for patient care. · FDA Class II
Units
617 generators
Sold at
Worldwide Distribution - USA (nationwide) and the countries of Argentina, Australia, Austria, Belgium, Canada, Chile, Colombia, Czech Republic, Denmark, France, Germany, Greece, Hungary, India, Israel, Jordan, Netherlands, Norway, Poland, Portugal, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Taiwan, Venezuela and UK.
Description
Potential failure of HALO Energy Generator to enter "stand by" mode when initially powered on; and therefore can not be used for patient care.
Product
HALO Flex Energy Generators Manufactured by B¿RRX Medical (now Covidien) 540 Oakmead Parkway Sunnyvale, CA 95085. The HALO Flex Energy Generator is indicated for use for the coagulation of soft tissue. The HALO Flex System is indicated for use in the coagulation of bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to, the esophagus, including Barrett's Esophagus.
Manufacturer(s)
Covidien LLC
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →