RecallWatchUSA

Home / Recalls / Medical Devices

CareFusion 303, Inc. recalls BD Pyxis MedBank Cabinet Software v2.1.2.9 Legacy MedBank Cabinets with Matrix Drawers include: . CUBEX MedFlex op

Recalled October 5, 2022 · FDA recall Z-1785-2022 · CareFusion 303

Hazard
The automated dispensing cabinet software update included a .dll file that is not compatible with MedBank Cabinet stations with MedCart Matrix drawers. The cabinet system will stop functioning (Matrix drawers will stop opening in the MedBank Application) leading to delay in access to medications or supplies needed in residential or procedural care facilities. · FDA Class II
Units
686 cabinets
Sold at
U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS,KY, LA, MA, MD, ME,. MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY.

Description

The automated dispensing cabinet software update included a .dll file that is not compatible with MedBank Cabinet stations with MedCart Matrix drawers. The cabinet system will stop functioning (Matrix drawers will stop opening in the MedBank Application) leading to delay in access to medications or supplies needed in residential or procedural care facilities.

Product
BD Pyxis MedBank Cabinet Software v2.1.2.9 Legacy MedBank Cabinets with Matrix Drawers include: . CUBEX MedFlex operating with Windows 7 . MedBank 200 Main operating with Windows 7 . MedBank 500 Main operating with Windows 7. An automated dispensing cabinet intended to securely store and dispense medications to health care providers in a clinical setting.
Manufacturer(s)
CareFusion 303, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov