RecallWatchUSA

Home / Recalls / Medical Devices

Boston Scientific Corporation recalls SpaceOAR Vue System, UPN SV-1010 and SV-2101. Used to temporarily position the anterior rectal wall away from the prost

Recalled May 4, 2022 · FDA recall Z-1003-2022 · Boston Scientific

Hazard
Users need to be aware that embolism is a possible adverse event associated with hydrogel misplacements. Updated IFU provides new procedural instructions, warnings and precautions that describe steps to verify correct placement of the hydrogel following implantation, as well as technique recommendations for the proper placement of SpaceOAR and SpaceOAR Vue. · FDA Class II
Units
23,368 units
Sold at
Worldwide - US Nationwide distribution.

Description

Users need to be aware that embolism is a possible adverse event associated with hydrogel misplacements. Updated IFU provides new procedural instructions, warnings and precautions that describe steps to verify correct placement of the hydrogel following implantation, as well as technique recommendations for the proper placement of SpaceOAR and SpaceOAR Vue.

Product
SpaceOAR Vue System, UPN SV-1010 and SV-2101. Used to temporarily position the anterior rectal wall away from the prostate during radiotherapy for prostate cancer to reduce the radiation dose delivered to the anterior rectum.
Manufacturer(s)
Boston Scientific Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov