Hazard
Certain dilation and extension catheters are being recalled because these catheters may have missed the hydrophilic coating step, which creates a lubricious surface on the devices, during the manufacturing process. · FDA Class II
Units
14,145 devices total
Sold at
Nationwide including Washington D.C. and Puerto Rico. OUS to include Canada.
Description
Certain dilation and extension catheters are being recalled because these catheters may have missed the hydrophilic coating step, which creates a lubricious surface on the devices, during the manufacturing process.
Product
Rebel Monorail PtCr Coronary Stent System, REF / Catalog No. H7493927324300, Rebel MR OUS 3.0mm x 24mm. The Rebel Monorail (MR) Stent System is a balloon expandable Platinum Chromium (PtCr) Alloy stent pre-mounted on a Monorail Catheter.