Home / Recalls / Medical Devices
Boston Scientific Corporation recalls Model 3300 LATITUDE Programming System with installed Model 3892 ALTRUA, INSIGNIA I, NEXUS I Software Support Applicatio
Recalled November 11, 2020 · FDA recall Z-0319-2021 · Boston Scientific
Hazard
There is potential when a user changes an EGM trace channel with a Manual Pace Threshold test in progress that the newly modified trace channel may change to a flat line. · FDA Class II
Units
8 devices
Sold at
Distributed in US - AR, FL, IL, and TX
Description
There is potential when a user changes an EGM trace channel with a Manual Pace Threshold test in progress that the newly modified trace channel may change to a flat line.
Product
Model 3300 LATITUDE Programming System with installed Model 3892 ALTRUA, INSIGNIA I, NEXUS I Software Support Application
Manufacturer(s)
Boston Scientific Corporation
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →