Boston Scientific Corporation recalls EMBLEM MRI S-ICD Pulse Generator Model A219; Subcutaneous Implantable Cardioverter Defibrillator

Recalled August 30, 2023 · FDA recall Z-2442-2023 · Boston Scientific

Hazard
There is a potential for a rare interaction between the EMBLEM S-ICD and LATITUDE communicator, which may cause S-ICD sensing disablement for a 24-hour interval. · FDA Class II
Units
41,411 units
Sold at
Worldwide

Description

There is a potential for a rare interaction between the EMBLEM S-ICD and LATITUDE communicator, which may cause S-ICD sensing disablement for a 24-hour interval.

Product
EMBLEM MRI S-ICD Pulse Generator Model A219; Subcutaneous Implantable Cardioverter Defibrillator
Manufacturer(s)
Boston Scientific Corporation