Home / Recalls / Medical Devices
Boston Scientific Corporation recalls Boston Scientific Coyote OVER-THE-WIRE PTA Balloon Dilation Catheter, 4F, 1) 2.5mm x120mm, 150 cm, UPM/Model No. REF H
Recalled January 21, 2026 · FDA recall Z-1042-2026 · Boston Scientific
Description
Boston Scientific is initiating the removal of certain batches of Coyote" Over-the-Wire (OTW) Percutaneous Transluminal Angioplasty (PTA) Balloon Dilatation Catheter devices due to the potential for inflation/deflation difficulty during use in the peripheral vasculature. An internal investigation found that a manufacturing process variation resulted in catheters that may not properly inflate/deflate. This was limited to Coyote OTW batches manufactured from 12Sep2025 through 13Oct2025. The processing variation has since been addressed. No other material numbers or batches are impacted by this removal.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →