RecallWatchUSA

Home / Recalls / Medical Devices

Boston Scientific Corporation recalls Accolade pacemaker, Models: a) L300 b) L301 c) L310 d) L311 e) L321 f) L331

Recalled July 14, 2021 · FDA recall Z-2000-2021 · Boston Scientific

Hazard
Latent release of small amounts of hydrogen within the pacemaker may cause a low voltage capacitor to become electrically compromised over time resulting in accelerated battery depletion of the battery and associated progression of displayed battery depletion indicators. · FDA Class II
Units
70074 units
Sold at
Worldwide distribution.

Description

Latent release of small amounts of hydrogen within the pacemaker may cause a low voltage capacitor to become electrically compromised over time resulting in accelerated battery depletion of the battery and associated progression of displayed battery depletion indicators.

Product
Accolade pacemaker, Models: a) L300 b) L301 c) L310 d) L311 e) L321 f) L331
Manufacturer(s)
Boston Scientific Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Boston Scientific Corporation recalls Accolade pacemaker, Models: a) L300 b) L301 c) L310 d) L311 e) L321 f) L331 | RecallWatchUSA