Biomet, Inc. recalls JUGGERKNOT IN¿GUIDE PUNCH, 2.9MM. The JuggerKnot 2.9mm In-Guide Punch is intended to be used to prepare a 2.9mm hole

Recalled September 26, 2012 · FDA recall Z-2399-2012 · Biomet

Hazard
Investigation found that the current design of the product allows the press-fit of the knob/shaft interface to be overcome by the impaction force of a mallet, causing the knob to potentially unfasten and slide down the shaft. · FDA Class II
Units
62
Sold at
Nationwide Distribution-including the states of WA, IN, AK, PA, GA, SD, KS, NY, VA, AR, AL, OH, WI, MA, and CA.

Description

Investigation found that the current design of the product allows the press-fit of the knob/shaft interface to be overcome by the impaction force of a mallet, causing the knob to potentially unfasten and slide down the shaft.

Product
JUGGERKNOT IN¿GUIDE PUNCH, 2.9MM. The JuggerKnot 2.9mm In-Guide Punch is intended to be used to prepare a 2.9mm hole into bone in which the JuggerKnot 2.9mm Suture Anchor is to be placed. The JuggerKnot 2.9mm Suture Anchor is intended to be used to attach soft tissue to bone.
Manufacturer(s)
Biomet, Inc.